Remove A Quad Receiver Module; Add A Quad Receiver Module - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Replaceable parts
Front cover

Remove a quad receiver module

Add a quad receiver module

6-26
CAUTION
EQUIPMENT DAMAGE—Do not remove or install Quad Receiver
Modules with power applied.
1.
Remove the front cover as described above.
2.
Unseat the quad receiver module by pulling the two retaining clips away from the
module.
3.
Pull module straight out of chassis.
1.
Remove the front cover as described above.
2.
Position a quad receiver module with the protruding side (with label) facing the
right side of the chassis.
3.
Slide the protruding edge into the upper and lower chassis track.
4.
Keep the retaining clips fully extended until the outer edges meet the receiver
cage.
ApexPro™
frontcover
2001989-351A

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents