GE ApexPro TT Technical Manual page 61

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

2001989-351A
7.
Proceed to the next digit by pressing the Graph button until the next LED
flashes.
8.
Repeat for all 4 (5 for T14 transmitter) digits.
The TTX number is not changed until the fourth/fifth digit is accepted by
pressing the Graph button. The top row LEDs flash twice upon acceptance of the
TTX number.
9.
To exit service mode, remove the short from pins 2 and 4. The service program
runs until the short is removed. Then the transmitter reloads the application code
and continues to run normally at the new TTX number.
ApexPro™
Installation and configuration
3-21