GE ApexPro TT Technical Manual page 56

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Installation and configuration
3-16
2.
At the PC, use a communication program such as HyperTerminal to set up the
ComPort connection.
Port:
Speed:
Parity:
Stop Bit:
Data Bits:
Flow Control:
3.
Press Enter to get the @> Enter Service Password: prompt.
4.
Type password, mms_aps (case-sensitive).
5.
Press Enter.
6.
At the @> prompt type ssf.
The command is "ssf x" where x identifies which antenna field(s) to enable.
Enter the "ssf x" command as listed below to configure the desired antenna
field(s).
X
Field D
1
Disabled
2
Disabled
3
Disabled
4
Disabled
5
Disabled
6
Disabled
7
Disabled
8
Enabled
9
Enabled
10
Enabled
11
Enabled
12
Enabled
13
Enabled
14
Enabled
15
Enabled
7.
Exit the communication program, then disconnect serial cable and PC, or
continue with step 6 in
ApexPro™
Com1 (comm port on the PC)
Baud 19200
No
1
8
Xon Xoff
Field C
Disabled
Disabled
Disabled
Enabled
Enabled
Enabled
Enabled
Disabled
Disabled
Disabled
Disabled
Enabled
Enabled
Enabled
Enabled
Setup the receiver on page
Field B
Field A
Disabled
Enabled
Enabled
Disabled
Enabled
Enabled
Disabled
Disabled
Disabled
Enabled
Enabled
Disabled
Enabled
Enabled
Disabled
Disabled
Disabled
Enabled
Enabled
Disabled
Enabled
Enabled
Disabled
Disabled
Disabled
Enabled
Enabled
Disabled
Enabled
Enabled
3-17.
2001989-351A