Guidance And Manufacturer's Declaration - Electromagnetic Immunity - GE ApexPro TT Technical Manual

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Guidance and manufacturer's declaration – electromagnetic
immunity
Immunity test
Electrostatic Discharge
(ESD)
EN 61000-4-2
Electrical Fast Transient/
Burst
EN 61000-4-4
Surge
EN 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
EN 61000-4-11
Power Frequency (50/60
Hz) Magnetic Field
EN 61000-4-8
2001989-351A
The ApexPro and CARESCAPE Telemetry Transmitter is intended for use in the
electromagnetic environment specified below. It is the responsibility of the customer
or user to assure that the ApexPro Telemetry Transmitter is used in such an
environment.
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
(>95% dip in U
) for 0.5
<5% U
t
t
cycles
<40% U
(>60% dip in U
) for 5
t
t
cycles
<70% U
(>30% dip in U
) for 25
t
t
cycles
<5% U
(>95% dip in U
) for 5 s
t
t
3 A/m
NOTE
U
is the AC mains voltage prior to application of the test level.
t
Compliance level
Electromagnetic environment – guidance
± 6 kV contact
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the
± 8 kV air
relative humidity should be at least 30%.
Not applicable
No AC input
Not applicable
No AC input
Not applicable
No AC input
3 A/m
Power frequency magnetic fields should be at
levels characteristic of a typical location in a typical
commercial or hospital environment.
ApexPro™
Electromagnetic compatibility
B-3