GE ApexPro TT Technical Manual page 186

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Electromagnetic compatibility
Electromagnetic immunity
Immunity test
Electrostatic Discharge
(ESD)
IEC 61000-4-2
Electrical Fast Transient/
Burst
IEC 61000-4-4
Surge
IEC 61000-4-5
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
Power Frequency (50/60
Hz) Magnetic Field
IEC 61000-4-8
B-8
The ApexPro receiver is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure that the
ApexPro receiver is used in such an environment.
IEC 60601 test level
± 6 kV contact
± 8 kV air
± 2 kV for power supply lines
± 1 kV for input/output lines
± 1 kV differential mode
± 2 kV common mode
<5% U
(>95% dip in U
) for 0.5
t
t
cycles
<40% U
(>60% dip in U
) for 5
t
t
cycles
<70% U
(>30% dip in U
) for 25
t
t
cycles
<5% U
(>95% dip in U
) for 5 s
t
t
3 A/m
NOTE
U
is the AC mains voltage prior to application of the test level.
t
Compliance level
± 6 kV contact
± 8 kV air
± 2 kV for power supply
lines
± 1 kV differential mode
± 2 kV common mode
<5% U
(>95% dip in U
)
t
t
for 0.5 cycles
<40% U
(>60% dip in U
)
t
t
for 5 cycles
<70% U
(>30% dip in U
)
t
t
for 25 cycles
<5% U
(>95% dip in U
)
t
t
for 5 s
3 A/m
ApexPro™
Electromagnetic environment – guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment.
Mains power should be that of a typical
commercial or hospital environment. If the
user of the equipment requires continued
operation during power mains interruptions,
it is recommended that the equipment be
powered from an uninterruptable power
supply or a battery.
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.
2001989-351A