GE ApexPro TT Technical Manual page 153

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

2001989-351A
1.
Turn on the RF monitor and allow it to operate for at least one hour for
temperature stabilization.
2.
Set the center frequency to the frequency of the transmitter. Refer to the ApexPro
Telemetry Frequency Chart Reference Manual.
a.
Press FREQ.
b.
Press the numbers corresponding to the frequency of the transmitter.
c.
Press MHz and press Enter.
3.
Set the span to 500KHz (50KHz/div).
a.
Press Span.
b.
Press 500.
c.
Press KHz and press Enter.
4.
Set the reference amplitude to -10dBm at 10dB/div.
a.
Press AMPT.
b.
Press 10.
c.
Press GHz/-dBm and press Enter.
5.
Connect the transmitter with lead wires to a patient simulator.
6.
Wrap the leadwires around the antenna of the RF monitor.
7.
Verify that a peak signal of at least -20dBm (-10dBm for T14 transmitter) can be
obtained. It may be necessary to move the transmitter and leadwires around to
obtain this peak signal.
ApexPro™
Checkout
316A
7-15