GE ApexPro TT Technical Manual page 187

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Note 1: At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by reflection from structures, objects, and
people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radio, AM and FM
radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the equipment
is used exceeds the applicable RF compliance level above, the equipment should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the equipment.
b
Over the frequency range 150 KHz to 80 MHz, field strengths should be less than 3 V/m.
2001989-351A
Guidance and Manufacturer's declaration - electromagnetic
immunity
The ApexPro receiver is intended for use in the electromagnetic environment
specified below. It is the responsibility of the customer or user to assure that the
ApexPro receiver is used in such an environment.
IEC 60601 test level
3 Vrms
150 KHz to 80 MHz
3 V/m
80 MHz to 2.5 GHz
ApexPro™
Compliance level
Electromagnetic environment – guidance
Portable and mobile RF communications
equipment should not be used closer to any part of
the equipment, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance
3 V rms
d = 1.2?
3 V/m
d = 1.2
d = 2.3
where P is the maximum output power rating of the
transmitter in watts (W) according to the
transmitter manufacturer, and d is the
recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
should be less than the compliance level in each
frequency range
Interference may occur in the vicinity of equipment
marked with the following symbol:
Electromagnetic compatibility
80 MHz to 800 MHz
800 MHz to 2.5 GHz
b
.
a
,
B-9

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents