GE ApexPro TT Technical Manual page 104

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Troubleshooting
5-20
NOTE
"XXX" is representative of the number of events stored in the error log. It
will appear on screen as an actual number.
7.
At the @> prompt type del XXX (where XXX equals the number of events as
indicated in the error message) and press Enter to display the most recent event.
Other commands include:
Type , (comma) and press Enter to display the previous log entry.
Type . (period) and press Enter to display the next log entry.
Type dael and press Enter to dump all entries. This is useful in conjunction with
the "capture-to-file" feature of Hyperterminal.
The following is an example of the information in each event log.
Record Number 244
Date/Time:
Error Code:
Severity:
Error Text:
Process Name:
Status Register:
User Stack Pointer:
Program Counter:
Super Stack Pointer:
Heap Pointer:
Passed Parameter:
Error Number:
Ticks:
The date and time are set by the ApexPro Telemetry System. If the date and time have
not been updated by the time the event occurs, the time defaults to 1 January 1990
00:00.
The error text explains the error code.
The severity levels are:
4 = An event occurred, not an error
5 = A minor error
6 = An error that caused the system to halt
7 = An error that caused the system to reset
The number of ticks indicates when the event occurred relative to the last time the
system was rebooted. There are 60 ticks in one second.
ApexPro™
1 January 2000 00:00
0x438a
4
Set fields for rcvr 15 to 15
sysstart
0x9042
0xff0902c8
0xff013a24
0xffcbffe8
0xff0c8554
15
60544
525
2001989-351A