GE ApexPro TT Technical Manual page 145

Designed to acquire and monitor physiological data for ambulating adult and pediatric patients within a defined coverage area.
Table of Contents

Advertisement

Electrical safety tests
2001989-351A
2.
Press Enter and minimize the MS-DOS screen.
3.
At a CIC, right-click on an empty available bed. Select newly admitted test bed
and verify ECG waveform appears without dropout.
4.
Restore the MS-DOS window either at the CIC, or return to your ApexPro
Telemetry System and restore the MS-DOS window there.
5.
Type discharge <unit>|<bed> and press Enter.
6.
At the message, Are you sure you want to delete?, type y.
7.
Repeat these steps for all remaining receivers.
Electrical safety tests provide a method of determining if potential electrical health
hazards to the patient or operator of the device exist.
Recommendations
Test conditions
Electrical safety tests may be performed under normal ambient conditions of
temperature, humidity, and pressure.
Test equipment
The recommended test equipment required to perform electrical safety tests is listed
below.
Item
Leakage Current Tester
Digital Multimeter (DMM)
Ground Bond Tester
ECG Test Body
Power outlet test
Before starting the tests, the power outlet from which the monitoring device will get
electrical power must be checked. This test checks the condition of the power outlet
to ensure correct results from leakage tests.
For international power outlets, refer to the internal standards agencies of that
particular country. Use a digital multimeter to ensure the power outlet is wired
properly.
If other than normal polarity and ground is indicated, corrective action must be taken
before proceeding. The results of the following tests will be meaningless unless a
properly wired power outlet is used.
ApexPro™
Checkout
Specification
Equivalent to the circuits shown
AC volts, ohms
0 – 1 ohm
All leads together
7-7

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents