Fluke Hydra Series II User Manual page 72

Data bucket
Table of Contents

Advertisement

2635A
Users Manual
ALRM
ENTER
ENTER
ENTER
ENTER
ENTER
2-22
Selecting Alarm 1 or Alarm 2.
LIMIT
(Alarm) key to access the alarm selection menu.
Use the up/down arrow keys to select alarm 1 or
1
2, then press the ENTER key.
2
Selecting the Alarm Mode. Press the up/down
LIMIT
arrow keys to select an alarm mode, OFF (Off), HI
OFF
(High), or LO (Low), then press the ENTER key.
HI
LO
Selecting the Alarm Numerical Value. Press
LIMIT
the up/down and left/right arrow keys to enter a
± 0 0 0 0 0
five digit number that defines the numerical value
- to -
for the alarm, ignoring the decimal point or scale
± 9 9 9 9 9
multiplier. The column being configured will have
a bright display. For example, for an alarm of
132.75V ac, enter +13275. When the entry is
correct, press the ENTER key.
Selecting the Alarm Decimal Value. Press the
LIMIT
left/right arrow keys to position the decimal point
.
in the number selected in the previous step. For
X
X X X X
.
the example above, the settings would be 132.75.
X X
X X X
.
When the decimal point is correct, press the
X X X
X X
.
ENTER key.
X X X X
X
Selecting the Alarm Scale Multiplier. Press the
LIMIT
up/down arrow keys to select the desired
x1
multiplier: x1 (x1), m (x .001), M (x1,000,000), or
m
k (x1,000). For the example above, the 132.75
M
multiplier would be x1.
k
correct, press the ENTER key.
configured, the procedure continues for alarm 2.
If alarm 2 is configured, the procedure terminates.
Typical
Required
Figure 2-13. Setting the Alarms
Press the ALRM
When the multiplier is
If alarm 1 is
op25f.eps