Fluke Hydra Series II User Manual page 65

Data bucket
Table of Contents

Advertisement

SET FUNC
FUNC
OFF
°F [°C]*
Hz
VAC
V DC
ENTER
SET FUNC
J
K
E
T
N
R
S
ENTER
b
C
Pt**
SET FUNC
2T
4T**
ENTER
0 0 0 . 0 0
- to -
9 9 9 . 9 9
ENTER
Figure 2-9. Configuring a Channel to Measure Temperature (RTDs)
Selecting the Temperature Mode.
FUNC key to access the SET FUNC (Set
Function) menu. Press the up/down arrow keys
until °C (Centigrade) or °F (Fahrenheit) is
displayed, then press the ENTER key.
opposite temperature scale is desired, refer to
Figure 2-2.)
Selecting the RTD Mode. Select Pt (Platinum),
then press the ENTER key.
Selecting the Terminal Mode. Select the two-
terminal (2T) or four-terminal (4T) mode.
automatically clears configuration data from the
channel a decade higher (n+10). For example,
selecting channel 3 for 4T operation will
automatically clear channel 13 of all configuration
data and lock it out from further use.
Selecting the "Ice Point" (R0).
Ro
up/down and left/right arrow keys to enter the
desired number, then press the ENTER key. The
default and most common setting is 100.00
(R0=100.00 ohms at 0 °C [32 °F]).
* Depends on
Typical
temperature scale.
Required
** Channels 1 to 10.
Front Panel Operations
Configuring a Measurement Channel
Press the
(If the
4T
Use the
2
op21f.eps
2-15

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2635a

Table of Contents