Using The Review Mode - Fluke Hydra Series II User Manual

Data bucket
Table of Contents

Advertisement

2635A
Users Manual

Using the Review Mode

Perform the procedure in Figure 2-18 to operate the Review Mode of operation. The
Review Mode is used any time during or after operation of the Scan Mode. While the
instrument is in the Scan Mode, the last, maximum, and measurements for each scanned
channel are stored in memory and updated with each scan, forming the Review Array.
For example, during scan operations, N can be used to monitor the maximum
measurement of a channel in real time. The Review Array is cleared by a control
sequence (see Figure 2-18 below), or by changing any parameter of any channel or the
measurement rate. The Review Array can be printed out using the L key (see Figure
5-4). If the instrument is in the Review Mode and scanning using the memory card, any
illegal memory card operations is noted only with a double "beep." If a double beep is
heard, exit the Review Mode and check the memory card error (see Figure 2-16).
REVIEW
REVIEW
REVIEW
SHIFT
REVIEW
REVIEW
2-30
Examining the Review Values. Press the
REVIEW
REVIEW key to start the Review Mode. Use the
up/down arrow keys to select the channel for
CH
review, then use the left/right arrow keys to view
the LAST (Last), MIN (Minimum), and MAX
(Maximum) values recorded during the past scan
intervals (scan can be active or inactive). Repeat
REVIEW
for each channel of interest. A series of dashes,
- - - - -, indicates all review data has been cleared
LAST
either by a control sequence (below), or by
MIN
changing any parameter of any channel.
MAX
indicates an overload. Press the REVIEW key to
exit the Review Mode.
values will remain in memory for update with the
next scan interval.
Clearing the Review Array. Press the Review
REVIEW
key to start the Review Mode. Press the SHIFT
key and then the REVIEW key to clear all review
- - - - -
data from all channels. If the scan mode is not
active, the display will change to dashes (- - - - -).
If the scan mode is active, new values will
appear. Press the REVIEW key again to exit the
Review Mode.
Figure 2-18. Using the Review Mode
OL
If not cleared, review
op30f.eps

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2635a