External Trigger Input Test; Calibration - Fluke Hydra Series II User Manual

Data bucket
Table of Contents

Advertisement

External Trigger Input Test

The External Trigger Input Test verifies that the rear panel trigger input of Hydra Series
II is functioning properly.
1. Switch OFF power to the instrument and disconnect all high voltage inputs.
2. Remove the Alarm Output eight terminal connector module from the rear of Hydra
Series II and all external connections to it. Connect short wires (to be used as test
leads) to the GROUND and TR terminals. Leave other ends of wires unconnected at
this time. Reinstall the connector. Refer to Figure 7-5.
3. Switch power ON.
4. On Hydra Series II, select the VDC function, 30V range for channels 0 through
5.Select a scan interval of 30 seconds.
5. Select trigger ON to enable the external trigger input. Press the K and M
(TRIGS) buttons (the display shows TRIG), then press either the up or down arrow
buttons to display ON. Press E.
6. Press Hydra Series II's Q button. Hydra Series II should scan channels 0 through 5
once every 30 seconds.
7. During the interval when scanning is not occurring, connect (short) the test leads of
the TR and ground Alarm Output terminals.
Ensure the connection causes a single scan to occur.
8. Disconnect (open) the TR and ground connection.
Ensure the scan continues to execute at its specified interval.
+ –
+ –
9-16 V
DC PWR

Calibration

Refer to the Fluke Hydra Series II Service Manual (P/N 688868) for
calibration procedures. The instrument must be stabilized in an
environment with an ambient temperature of 22 to 24ºC and a relative
humidity of less than 70% and must have been turned on for at least 30
minutes prior to calibration.
The instrument features closed-case calibration controlled over the Computer Interface.
Using known reference sources, closed-case calibration has many advantages. There are
ALARM OUTPUTS
0
1 2 3 TR
0 1 2 3 4 5 6 7
0 1 2 3 4 5 6 7
0
1 2 3 TR
+30V
!
Figure 7-5. External Trigger Test
Note
DIGITAL I/O
Σ
Σ
Maintenance
7

Calibration

op67f.eps
7-21

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

2635a

Table of Contents