Fluke Hydra Series II User Manual page 21

Data bucket
Table of Contents

Advertisement

D
ALRM
SET ALARM
1
2
CANCL
Mx+B
SET Mx+B
M
+001.00
CANCL
SHIFT
RAtE
FASt
SLO
CANCL
E
Selecting Alarms. Press the ALRM key to open
the alarm menu. Each configured channel can
have two alarm limits assigned. An alarm is set
when a reading is below or above an alarm limit.
Configuration starts with an alarm limit selection,
1 or 2. Press CANCL to exit. [Figure 2-13]
Setting Mx+B Scaling. Press the Mx+B key to
open the Mx+B menu.
arrow keys select the digits for the first parameter
(M) (default +001.00). The effect of Mx+B scaling
is to take a measurement (x) and modify it by
multiplying the measurement with M and then
adding an offset B (configured after M is set). For
example,
Mx+B=+1.5x+25
measurement
1.5(20.000) + 25 = 55.000. Press CANCL to exit.
[Figure 2-14]
Selecting the Measurement Rate.
SHIFT key, release, then press the Right Arrow
key to open the RATE menu.
measurement portion of the scan interval, the
measurement rate can be FASt (Fast) or SLO
(Slow).
The slow rate gives full 5-digit
measurement resolution, while the fast rate gives
only 4-digit resolution. The advantage of a fast
measurement rate is more readings during
continuous scanning or low scan intervals. Press
CANCL to exit. [Figure 2-12]
Ten Minute Tour (cont)
xix
Ten Minute Tour
Up/down and left/right
applied
to
of
20.000
would
display
Press the
During the
(continued)
a
op79_5f.eps