2635A
Users Manual
F-2
Offset
Description
=====
===========
0
File type; 0 for setup files, 1 for data files.
1
File format; always zero for this format.
2-81
Tag string.
Copied from the setup file used with a time
string appended; or just a time string if no setup file used.
82
Set up format version; always zero.
83
Global instrument configuration; 0x01 bit set for degrees
Fahrenheit, else Centigrade; 0x02 bit set for open
thermocouple checking enabled; 0x80 bit set for alarms on open
thermocouple. See TEMP_CONFIG.
84
Measurement rate; 0 for slow, 1 for fast.
85
Trigger type; 0 for off, 1 for on, 2 for monitor alarm.
TRIGGER.
86
Output format type; 1 for no units, 2 for units.
87
Totalizer debounce; 0 for off, 1 for on.
88
Interval BCD hours.
89
Interval BCD minutes.
90
Interval BCD seconds.
91
Event Status Register (ESR) value.
92
Service Request Enable (ESE) register value.
93
Instrument Event Enable (IEE) register value.
94
Logging status; 0x1 bit set if logging enabled, 0x02 bit
set if LOG queue stops logging when full.
95
Logging filter; 0 for all, 1 for alarms, 2 for alarm
transitions.
96
Logging destination; 0x01 bit for printer, 0x02 bit set for
LOG queue, 0x04 bit set for memory card.
97
Front panel lock mode; 0 for not locked, 3 for
configuration lock.
98-727
Channel definitions:
Offset within each channel definition
Description
———
———------——
0
Measurement. 0 if off, 1 if vdc, 2 if vac, 3 if
ohms, 4 if frequency, 9 if thermocouple, 11 if RTD.
1
Measurement range.
range, with 0 being the lowest range for the
measurement type (e.g., 0 for 300 mVDC).
0x10 bit set if autoranging.
2
Thermocouple/RTD type.
3 for E, 4 for T, 5 for N, 6 for R, 7 for S, 8 for
B, 9 for C.
3
Alarm flags.
bit set for SP1 high, 0x04 bit set for limit 2 (SP2)
low, 0x08 bit set for SP2 high.
4-7
Alarm limit 1 (4-byte single-precision floating point).
8-11
Alarm limit 2 (4-byte single-precision floating point).
12
Alarm 1 digital I/O association.
13
Alarm 1 display range.
14
Alarm 2 digital I/O association.
15
Alarm 2 display range.
16-19
Mx+B M value (floating point).
20-23
Mx+B B value (floating point).
24
Mx+B M display range.
25
MX+B B display range.
26-29
RTD R0 value (floating point).
728-729 CRC-16 of the data from offset 82 to 727.
See INTERVAL.
See PRINT_TYPE.
See LOCK.
Lower bit values are measurement
0 for Pt, 1 for J, 2 for K,
0x01 bit set for limit 1 (SP1) low, 0x02
See SCALE_MB (DISPLAY CODE - 1).
See SCALE_MB (DISPLAY CODE - 1).
See SCALE_MB (DISPLAY CODE - 1).
See SCALE_MB (DISPLAY CODE - 1).
See RATE.
See
See FORMAT.
See TOTAL_DBNC.
See LOG_MODE.
See PRINT_TYPE.
See RANGE?
See ALARM_LIMIT.
See ALARM_ASSOC.
See ALARM_ASSOC.
See RTD_R0.