Table 7-5. Performance Tests for RTD Temperature Function (Resistance Source) (DIN/IEC 751
DECADE RESISTANCE
SOURCE
These figures assume that RTD R0 is set to 100.00 ohms for each channel.
Accuracy given is for 4-wire measurements only. For 2-wire measurements, degrade the accuracy
specifications by 5.2 °C per ohm of single lead wire resistance. For 2-wire measurements, degrade the
accuracy by an additional 11°C (channels 1 .. 20)or 0.05°C (channel 0).

RTD Temperature Accuracy Test (Using DIN/IEC 751 RTD)

1. Switch OFF power to the instrument and disconnect all other high voltage inputs.
2. Remove the Input Module from the rear of the instrument. Open the Input Module
7-14
11
SOURCE
H L
(4-WIRE)
SENSE
H L
(4-WIRE)
1
Figure 7-3. Four-Terminal Connections to Decade Resistance Box
TEMPERATURE SIMULATED
100 e
200 e
300 e
and connect a Platinum RTD, conforming to the European Standards IEC 751 (DIN
43760).
Two-terminal RTD: Connect the RTD's excitation leads to the H (high) and L (low)
terminals of channel 1.
Four-terminal RTD: Connect the RTD's excitation leads (one red and one black
wire) to the H (high) and L (low) terminals of channel 1. Connect the RTD's second
pair of red and black leads to the H and L leads of channel 11. (Refer to Figure 7-3
for proper connection.) Reinstall the Input Module into the instrument.
4-WIRE (4T) CONNECTION
12
13 14 15 16 17 18 19 20
H L H L H L H L H L H L H L H L H L
H L H L H L H L H L H L H L H L H L
2
3
4
5
6
7
DECADE
RESISTANCE
BOX
Amendment 2) (ITS-90)
°C
0
266.35
557.69
HYDRA
INPUT
MODULE
8
9
10
TEMPERATURE ACCURACY
1 YEAR @ 18-28°C
± 0.12°C
± 0.22°C
± 0.37°C
op65f.eps