2020 Software Manual
Calibration Using the Flow-Match
Regulator
Connecting the Flow-Match Regulator to
the Cylinder
WARNING
To connect the flow-match regulator to the gas cylinder:
70
Observe proper handling procedure for all gases! See Compressed
Gases on page 25.
1. Connect the regulator to the calibration gas cylinder.
If you are using a portable tank of calibration gas (Part
No. MX350012), connect the regulator (Part No.
MX350006) directly to the tank.
2. When the regulator is connected properly, you can read
the cylinder contents from the regulator gauge.
3. Connect the adapter tubing to the regulator.
2020 Sample
Figure 11
Open/Close Valve
Flow Rate
Indicator
Regulator
Adapter Tubing
Contents
Gauge
Probe
Calibration
Gas Cylinder
Flow-Match Regulator
Regulator