Support Equipment and Consumables

Calibration

For normal operation these items are required:
28
4. AC Adapter or AC Adapter with AC Line
Cord
5. Wrist Strap
6. Replacement Sample Inlet Filters (10 pieces)
7. Reference Card
1. Calibration Gas Regulator (Part No. MX350006).
2. Calibration gas containing 100 ppm isobutylene. (Part
No. MX350012). Other concentrations of the calibration
gas may be required. This will depend on your
application.
3. Zero air. Occasionally, clean, ambient air is suitable for
calibration.
Alternatively, you can use a gas sampling bag and a
source of hydrocarbon free air. Air should not have
more than 0.1 ppm total hydrocarbons (THC).
If you will be using large tanks of gas, specify a single
stage, high purity regulator with a CGA 590 connection
at the inlet. The regulator should also have a 1/8"
parallel, compression fitting with which to connect the
regulator to the gas bag adapter. The delivery pressure
must be adjustable to between 5 and 20 psig (34.5 and
138 kPa). You may require a gas bag and gas bag