2. Select "Set", "Cal" and "Mem".
3. Select the desired Cal Memory (1 to 15) with the
4. Press "Chng" to change the parameters of the Cal
5. If you selected "User", enter the name, response factor
6. If you entered "Lib", use the "Next" and "Prev" keys to
NOTE: It does not matter which Cal Memory is selected or which
response factor is entered, 2020's response is not specific to
any one compound. The reading displayed represents the total
concentration of all ionizable compounds in the sample.
7. Calibrate the instrument as described in Calibration
8. Repeat this procedure for each Cal Memory you need.
Whenever the instrument is calibrated, 2020 updates the selected
Cal Memory only. Each Cal Memory must be calibrated at least
once a day. Frequency of calibration will depend on ambient
conditions and instrument response. If ambient conditions change or
the response has drifted, a calibration must be performed for each
Cal Memory to ensure reliable operation.
all the Cal Memories, however, you must refill the bag
before zeroing each Cal Memory.
"Next" and "Prev" keys.
Memory. Select "User" or "Lib".
and alarm levels.
select the required library. See page 174 for a list of
Library entries.
Using the Flow-Match Regulator on page 70. When
the calibration is completed, the calibration information
is automatically stored in the selected Cal Memory.
Detailed Operation
77