Maintenance of the UV Lamp

Removing and Replacing the UV Lamp

NOTE: Do not remove or replace the UV lamp in a hazardous location.
To remove the UV lamp from your instrument:
WARNING
WARNING
116
1. Ensure the instrument is turned off.
You must turn the instrument off before removing the lamp cover.
2. Use the 2020 multi-tool to remove the lamp housing
cover.
3. Tilt 2020 slightly and remove the UV lamp.
Do not touch the wire grid inside the detector cell. Any dust or dirt in
the detector cell can be blown out with a gentle jet of compressed air.
Do not insert any object, other than the UV lamp, into the lampholder.