2020 Software Manual
NOTE: It does not matter which Cal Memory is selected or which
response factor is entered. 2020's response is not specific to
any one compound. The reading displayed represents the total
concentration of all ionizable compounds in the sample.
When using response factors, results are expected to be accurate to
+/- 10 ppm or +/- 25%, whichever is greater.
Compound
Acetaldehyde
Acetone
Acetophenone
Acrolein (2-Propenal)
Acrylic Acid
Acrylonitrile
Allyl Chloride (3-Chloro-1-Propene)
Benzene
Bromoform (Tribromomethane)
1,3-Butadiene
n-Butane
n-Butanol
n-Butyl Acetate
Butyl Acrylate
Butyl Cellosolve
n-Butyl Acrylate
n-Butyl Mercaptan (Butanethiol)
Carbon Disulfide
Chlorobenzene
170
Response
Factor
10.5
1.2
2.0
4.0
10.9
ND
3.9
0.5
2.0
0.7
0.2
3.4
2.3
1.8
3.1
1.8
0.6
1.3