Setting The Cell Constant Manually - wtw inoLab Cond Level 1 Operating Manual

Laboratory conductivity meter
Hide thumbs Also See for inoLab Cond Level 1:
Table of Contents

Advertisement

Setting the fixed cell
constant
0.010 cm

4.3.2 Setting the cell constant manually

Note
The cell constant to be set up must either be taken from the
operating manual of the measuring cell or is printed on the
measuring cell.
You can set the the cell constant manually as follows:
1 Press the
the display.
-1
2 Press the
3 Press the
0.010 cm
4 To return to the measuring mode: Press the
From now on, the cell constant 0.010 cm
used.
c
key repeatedly until CELL appears on
g
key.
c
key repeatedly until the cell constant
-1
appears on the display.
Operation
m
key.
-1
will be
25