wtw inoLab Cond Level 1 Operating Manual page 23

Laboratory conductivity meter
Hide thumbs Also See for inoLab Cond Level 1:
Table of Contents

Advertisement

Note
If error message E3 appears see chapter 6 W
AutoRead
During calibrating, the AutoRead function is automatically
activated. The AR display flashes. The calibration procedure
is finished when the AR display stops flashing.
4 Immerse the measuring cell into the 0.01 mol/l KCL
control standard.
g
5 Press the
– If no temperature probe is connected, enter the
current temperature of the solution using
and confirm with
– If a temperature probe is connected, the AR
measurement to determine the cell constant starts.
AR flashes until a stable signal is achieved. The
determined cell constant ist displayed; the measuring
instrument automatically stores the cell constant.
key.
g
.
Operation
u d
...
HAT TO DO IF
23