4 Programming Reference

Calibration Commands

Read the
calibration
accuracy and reliability.
CALibration:DATE <"date">
CALibration:DATE?
Stores the calibration date in nonvolatile memory. Enter any ASCII string up to 16 characters. The
query returns the date. If no date is stored, an empty quoted string ("") is returned.
Parameter
<"date">String program data. Enclose string parameters in
single or double quotes.
Enter a calibration date: CAL:DATE "2/2/2017"
CALibration:END
Ends the calibration process. This command should be used after the calibration constants have been
set, but before storing the calibration using CAL:STOR.
Parameter
(none)
End calibration: CAL:END
CALibration:LOAD
Loads the previous calibration values. For example, this can be used if there has been an error in the
calibration procedure. This command is only valid prior to executing the CALibration:STORe command.
Once the CALibration:STORe command has been executed, the previous calibration values are
overwritten.
Parameter
(none)
Loads the previous calibration values: CAL:LOAD
CALibration:RESult? <value>
Used to enter the resulting measurement for calibration. The value entered depends on the
calibration step. The command returns a zero (0) if the calibration step was successful.
52