Usb Connections; Digital Port Connections; Thermistor Connections - Keysight Technologies BT2152A Operating And Service Manual

Keysight bt2152a self-discharge analyzer
Hide thumbs Also See for BT2152A:
Table of Contents

Advertisement

2 Installing the Analyzer

USB Connections

1. Connect the instrument to the USB port on your computer using a customer-supplied USB cable. .
2. With the Connection Expert utility of the Keysight IO Libraries Suite running, the computer will
automatically recognize the instrument. This may take several seconds. When the instrument is
recognized, your computer will display the VISA alias, IDN string, and VISA address. This information is
located in the USB folder.
3. You can now use Interactive IO within the Connection Expert to communicate with your instrument,
or program the instrument directly using the applicable

Digital Port Connections

Twist and shield all signal wires to and from the digital connector. With shielded wire, connect
only one end of the shield to chassis ground to prevent ground loops.
An 8-pin connector and a quick-disconnect connector plug are provided for accessing the digital port
functions. Refer to
Items Supplied
connections. The connector plug accepts wires sizes from AWG 14 (1.5 mm 2 ) to AWG 28 (0.14 mm 2 ).
Strip the wire insulation back 7 mm.
1. Insert wires
2. Tighten screws
3. Fault/Inhibit configurable pins
(observe INH polarity)
4. Digital IO-configurable pins
5. Signal common
Information on using the digital port is found under
are described under
Common

Thermistor Connections

A 2-pin connector and a quick-disconnect connector plug are provided for the thermistor cable
connections. Use the provided thermistor cable. Refer to
number.
1. Insert wires
2. Tighten screws
The thermistor provided with the cable is Keysight p/n 0837-4194; Amphenol Advanced Sensors
RL1005-5744-103-SA. Refer to
28
for part numbers. Disconnect the connector plug to make your wire
Using the Digital
Characteristics.
SENse:THERmistor:CONFigure
SYSTem:COMMunicate:LAN
Port. The electrical characteristics
Items Supplied
for the thermistor cable part
for additional information.
Keysight BT2152A Operating and Service Guide
commands.

Advertisement

Table of Contents
loading

Table of Contents