Command string
:FQM:ENAB?
:FRZE:BUSY?
:INSP:BUSY?
:INSP:ERR?
:INSP:PRF:NAME
<STRING>
:INSP:PRF:NAME?
:INSP:PRF
:NAMES?
:INSP:TBL:SHOW
<BOOLEAN>
:INSP:TBL:SHOW?
:INSP:TIP:NAME
<STRING>
:INSP:TIP:NAME?
OLP-88
Parameter type / Response type / Unit / Info
Returns TRUE if the calculated focus quality metric for video frames is
enabled.
Response type:
<BOOLEAN>
See :FQM:ENAB command for more information.
Returns TRUE if an image is captured. After saving the image, the value
is changed to FALSE.
Response type:
<BOOLEAN>
Returns TRUE if the analysis of an image has been started. After
analyzing the image, the value is changed to FALSE.
Response type:
<BOOLEAN>
Returns TRUE if the test of an image causes an error.
Response type:
<BOOLEAN>
Name of the selected inspection profile name. The first entry in the
common profile names directory is used as default name.
See :INSP:PRF:NAMES? command for more information.
Returns the name of the selected inspection profile.
Response type:
<STRING>
See :INSP:PRF:NAME command for more information.
Returns the names of all available inspection profiles (read from
common (PCM and MSCOPE) profile directory).
Shows or hides the result table for a tested image.
Default:
TRUE
Returns TRUE if the result table for a tested image is displayed.
See :INSP:TBL:SHOW command for more information.
Selects one tip out of the tip names map. See :INSP:TIP:NAMES?
command for more information.
Each time the tip names map changes, the tip name has to be updated.
Returns the name of the selected tip.
Response type:
<STRING>
See :INSP:TIP:NAME command for more information.