6 P
/PCM O
ROBE
PERATION
C
D
P
ONFIGURING THE
IGITAL
44
ROBE
Press the [A] key and then the
overlay view. Repeat the action until the desired view appears.
Three views can be displayed: without overlay, with colored
edges, and colored edges with a legend.
Overlay
A summary of test results is displayed at the upper right of the
screen.
Zone A: represents the core zone. It is
the area surrounding the core.
Zone B: represents the cladding zone. It
surrounds the majority of the fiber
cladding.
Zone C: represents the epoxy ring.
Zone D: represents the ferrule zone. It
identifies a portion of the ferrule near
and around the fiber.
Note: To return to a Live image, press the QuickCapture™ key
or press the [A] key and tap the
both the live image and a test result simultaneously.
If the test has been passed, the image has a green colored
frame. If failed, the frame is red.
Fig. 18 Pass/Fail analysis – overlay view
Magnification Control
The Magnification Control allows you to modify the live display
from high to low magnification and vice-versa. In the high
magnification mode, automatic centering is available.
[Overlay]
button to change the
Zone D
Zone C
Zone B
Zone A