Pressure Units Setting; Access Code Settings - GE Druck TRX-II User Manual

Multi-function calibrator
Table of Contents

Advertisement

Changing the temperature scale
Leave the cursor at NONE at the measure and source menu and
press the # and EXE keys simultaneously to open the set-up menu.
Select Temp. settings and press EXE. Select °C and press EXE to
change from °C to °F or vice versa. Press CE to quit the set-up
menu.

Pressure Units Setting

Pressure readings can be shown in different pressure units. The
selection can be made via the Set-up menu.
Changing the pressure units
Leave the cursor at NONE at the measure and source menu and
press # and EXE keys simultaneously to open the set-up menu.
Select PRESSURE UNITS and press EXE. Select one of the 10
different pressure units and press EXE to change. Press CE to leave
the set-up menu.
Note: The reference temperature for in H
changes according to the defined temperature setting °C or °F.
°C refers to water temperature at 4°C and °F to 68°F.

Access Code Settings

The ACCESS CODE is programmable by the customer. The
following menus are protected by this code;
ACCESS CODE.
Calibration menu (Electrical).
Add pressure sensors.
Delete pressure sensors.
Calibration pressure sensors.
Access Code 9410 is the standard factory setting.
K319 Issue 2
0 or mH
0 automatically
2
2
EN-9