Before starting the calibration, check the time and date on the
instrument are correct.
To do a calibration on a measure or source function, use the
advanced menu option.
Select:
Enter the calibration PIN: 4321
Select the
Select:
Then select a function and start the calibration:
1. Select channel
2. Select function
Page 69 of 79
Table 5-1
Calibration equipment
(ppm = parts per million)
Current (mA) calibrator.
Accuracy - Current measure/source,
See Table 5-2 and Table 5-3.
Volts calibrator.
Accuracy - Voltage measure/source,
See
or Table 5-6.
Table
5-5
mV calibrator.
Accuracy - Millivolts measure/source,
See Table
.
5-4
All Ranges total uncertainty of 0.025% reading or
better.
UPM only. Refer to the user manual for the IDOS UPM.
DASHBOARD >>
>> ADVANCED
button.
PERFORM CALIBRATION
SETTINGS
[EN] English – K0571 Issue 1
Druck DPI611