GE Druck DPI611 User Manual page 73

Portable pressure calibrator
Hide thumbs Also See for Druck DPI611:
Table of Contents

Advertisement

User Manual
Applied
mV
±2000
±1000
±200
±100
0 (short circuit)
Applied
V
±30
±21
±20
±10
±5
0
Page 73 of 79
Volts (V): 0, 5, 10, 20, 21, 30.
Check that the error is within limits (See Table 5-4
and Table 5-5).
Table 5-4 Millivolts (measure) error limits
Calibrator
uncertainty
(mV)
0.051
0.040
0.051
0.0040
0.0036
Table 5-5 Voltage (measure) error limits
Calibrator
uncertainty
0.00052
0.00040
0.00031
0.00016
0.00008
0.000024
(V)
[EN] English – K0571 Issue 1
Druck DPI611
Permitted
DPI611 error
(mV)
0.1280
0.0940
0.0148
0.0110
0.0070
Permitted
DPI611 error
(V)
0.00180
0.00154
0.00118
0.00089
0.00075
0.00060