User Manual
LINEARITY -
TOLERANCE -
4.3
ANALYSIS FUNCTION
Set Input channel parameters (See Section 4.2), and return to
CHANNEL SETUP screen.
Select the Start button
The Analysis window displays the following:
The deviation of each Input channel from the ideal transfer
•
characteristic.
A tolerance limit test icon
•
To check the full range of the device do the following:
1. Step the Reference signal value through its range.
2. Check the Analysis window at each step.
3. If the Reference is sourced from the calibrator, move to the
channel window to change the Reference value.
4. Return to the Analysis window.
When the analysis is complete exit the window by selecting
5.
exit
Page 64 of 79
The transfer characteristic from Reference to Input
signal. This can be one of the following:
Linear - a proportional response.
Square Root - often found in flow sensors.
The test limits for the deviation from the transfer
characteristic.
.
Pass
Fail
.
(within tolerance test limit)
(outside tolerance test limit)