User Manual
Table 2-2
Step Procedure (Vacuum)
1
Set to vacuum operation
(-).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
3
Seal the system.
4
Use the pump to set the
maximum vacuum or set
the vacuum you want to
adjust.
5
Adjust the vacuum
(+ decrease; - increase).
2.2
CALIBRATOR OPERATIONS
2.2.1
Basic Calibrator Operation
1.
Select:
2.
Select the channel by performing the following tasks.
•
Page 31 of 79
DASHBOARD >>
Swipe to the TASK MENU by swiping the display
from right to left.
Table 2-3
Step
Procedure (Pressure)
1
Set to Pressure operation
(+).
2
To do equal adjustments
(up or down) at the end of
the procedure, turn the
volume adjuster to the
middle of its range of
operation.
3
Seal the system.
4
Use the pump to set the
maximum pressure or set
the pressure you want to
adjust.
5
Adjust the pressure
(+ decrease; - increase).
CALIBRATOR
[EN] English – K0571 Issue 1
Druck DPI611