IEC 60601 -
immunity tests
Test level
Conducted HF
3 V rms 150 kHz to 80 MHz
noise IEC 61000-
outside the ISM band, 6 V RMS
4-6
in the ISM and amateur radio
bands
Modulation:
80% AM at 1kHz
Radiated HF noise
10 V/m 80 MHz to 2,7GHz
IEC 61000-4-3
Modulation:
80% AM at 1kHz
27 V/m 385 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 450 MHz
Modulation:
FM + 5 Hz deviation: 1 kHz sine
9 V/m 710, 745, 780 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 810, 870, 930 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 1720, 1845 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 1970, 2450 MHz
Modulation:
Pulse Modulation: 217 Hz
9 V/m 5240, 5500, 5785 MHz
Modulation:
Pulse Modulation: 217 Hz
09/2018
1019179-00A
16.1.3
Table 204: Guidelines and manufacturer's
declaration – Electromagnetic interference
The beauty system has been designed for use in the electromagnetic envi-
ronment specified below. The device operator must ensure that it is used in
such an environment.
Defined
level
3 V rms 150 kHz to 80
MHz outside the ISM
band, 6 V RMS in the
ISM and amateur radio
bands
Modulation:
80% AM at 1kHz
10 V/m 80 MHz to 2,7GHz
Modulation:
80% AM at 1kHz
27 V/m 385 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 450 MHz
Modulation:
FM + 5 Hz deviation:
1 kHz sine
9 V/m 710, 745, 780 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 810, 870, 930 MHz
Modulation:
Pulse Modulation: 18 Hz
28 V/m 1720, 1845 MHz
Modulation:
Pulse Modulation: 217 Hz
28 V/m 1970, 2450 MHz
Modulation:
Pulse Modulation: 217 Hz
9 V/m 5240, 5500, 5785
MHzModulation:
Pulse Modulation: 217 Hz
Comment 1: The higher value applies to 80 MHz und 800 MHz.
Comment 2: These guidelines might not apply to all situations. The propagation of
electromagnetic waves is affected by absorption and reflexes caused by
buildings, objects and people.
a
The field strength of a stationary transmitter e.g. standard cell phone stations (mobile /
cordless phones) and public mobile radio services, amateur stations, MW and UHF radio
and TV transmitters) cannot be precisely determined for theoretical reasons. It is recom-
mended that an onsite test should be carried out in order to determine the electromagnet-
ic environment caused by stationary HF transmitters. If the field strength of the massage
device determined at the installation site exceeds the defined HF level listed above then
the device must be monitored with regard to its normal operation. If any abnormal charac-
teristics are recorded then it will be necessary to introduce additional measures, such as
reorientation or converting the device.
b
The field strength must be less than 3 V/m over the frequency ranges from 150 kHz to
.
80 MHz
Electromagnetic environment – Guidelines
Portable and mobile radios are not to be used at
shorter distances to the Model-No.: Model-No.:
JK 158 / BA 32-0 VS and its cables other than the
recommended safe distance, which has been
calculated according to the transmitting frequency
are made.
Recommended safe distance:
80 MHz to 800 MHz
800 MHz to 2,7 GHz
P is the transmitter's maximum output power
(nominal rating) in Watts (W) according to the de-
tails provided by the manufacturer of the transmitter
and the recommended safe distance is given in
metres (m).
The field strength of the stationary radio transmitter
is for all frequencies
in accordance with tests that
b
were carried out onsite
less than the determined
a
level.
Interference is possible in the immediate surround-
ing of equipment that carry the following symbols.
Operating instructions – 81/92
Appendix

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents