2.32 $AAS1

Description:
Reloads the factory calibration parameters.
Syntax:
$AAS1[CHKSUM](CR)
$
Delimiter character
AA
Address of the module to be reloaded (00 to FF)
S1
Command to reload the factory calibration
parameters
Response:
Valid command :
Invalid command: ?AA[CHKSUM](CR)
!
Delimiter character for a valid command
?
Delimiter character for an invalid command
AA
Address of the responding module (00 to FF)
There will be no response if the command syntax is
incorrect, there is a communication error, or there is no
module with the specified address.
116
I-7021/22/24, M-7022/24/28 User Manual, Rev: B2.3
!AA[CHKSUM](CR)