NURSING BEDS

12.5 Inspection of electrical components

Description
Visual inspection of electrical components:
mains cable for kinks and damage
hand set cabling for kinks and damage
mains plug for damage
internal cabling for kinks and damage
(especially at critical locations such as cable leadthroughs,
movable mattress support parts and the lifting device)
correct cable routing
Correct seating and condition of all:
plug contacts
plug contact sealing rings
cable leadthroughs
cable fixings
tension protection
anti-kink sheaths
Performance inspection of electrical components:
Motor switch-off caused by limit position switches
by running all adjustments up as far as the possible
limit position
Switch-off is working correctly if:
• the switching noise in the switch-off relay is audible
motors (noises, speed)
keys on remote control
(function, use of movement, damage)
Casa
Med Classic FS
Testing plan
Assessment
OK
defective
57
Issued: 25.07.2007