6-Monthly Inspection - Check Control Unit; Testing Intervals - Days Casa Med Classic FS Operating Manual

With dewert drive systems. nursing beds
Table of Contents

Advertisement

NURSING BEDS
12.2 6-monthly inspection - check
control unit
Preparation:
Lock all adjustment functions at the
hand set, then carry out performance
test as described in the following
"Check" section.
Check:
Test all functions by activating the
relevant function keys on the hand set.
None of the adjustment functions
should react.
If an adjustment functions correctly,
the control unit must be replaced.

12.3 Testing intervals

Testing intervals are dependant on the type of usage for Days
Healthcare nursing beds.
• Days Healthcare prescribes testing intervals of two years for safety
inspections in accordance with MPBetriebV (Germany) under normal
environmental and operating conditions.
• If being brought back into use, inspections should take place before
every new usage (after being assembled at the patient's location) and
then every two years.
• Testing intervals should be shortened if the nursing bed has been
subjected to higher loading (for example by frequent patient changes
in a nursing home). In case of doubt, please contact Days Healthcare
for advice.