Routing The Hand Set Cable; Carrying Out A Performance Test - Days Casa Med Classic FS Operating Manual

With dewert drive systems. nursing beds
Table of Contents

Advertisement

NURSING BEDS

7.11 Routing the hand set cable

NOTE
Tension relief devices (e) are provided
for on both sides of the nursing bed.
You should select the side to suit the
installation site.
Loosen the tension relief (e) for the hand
set cable (HB), insert the cable and
screw the tension relief tight.

7.12 Carrying out a performance test

NOTES
• Please note the information in chapters 9.2 and 9.4 "Operation"
when plugging in the mains plug and checking the adjustment
functions.
• Ensure that the transport safety devices at the head and foot-
pieces of the mattress support have been removed (Chapter 7.4).
• When operating the adjustment motors, ensure that the nursing
bed cabling cannot be crushed or overstretched.
Plug in the mains plug.
Check all nursing bed adjustments
for correct working order.
Run the nursing bed up to its highest
position.
Remove the mains plug from the socket.
Setting up the nursing bed
Casa
Med Classic FS
Tension protector
position (head end)
Cable hand set
e
Plug in the
mains plug
Issued: 25.07.2007
HB
29