Notes On Setting Up The Nursing Beds - Days Casa Med Classic FS Operating Manual

With dewert drive systems. nursing beds
Table of Contents

Advertisement

NURSING BEDS
Risk of short-circuit and fire!
• Before moving the nursing bed, unplug the mains cable and
ensure that the mains cable cannot be run over.
• The nursing bed may not be used in surroundings in which it is
possible that flammable gases or vapours
(e.g. from anaesthetics) are present.
Danger due to actuator motor overheating!
• Before adjusting the bed height, ensure that nothing is blocking
the full adjustment movement.
• Avoid using the actuator motors for more than two minutes.
Maximum continuous operation of actuator motors = 2 minutes.
Minimum pause after 2 minutes operation = 18 minutes.
• Never use more than two actuator motors simultaneously.
1.6

Notes on setting up the nursing beds

Injury hazard due to improper assembly!
• Ensure that all components in the nursing bed have been
correctly assembled.
• Check that all adjustment options are in perfect working
order after assembly.
Accident hazard!
• Only carry out assembly work using suitable tools.
• Only carry out assembly work if you are a competend
person who is used to working with the tools required.
Trapping hazard!
There is a possible danger of trapping at all joints on the nursing
bed and at support points for the mattress support. All hazard
points on the nursing bed are identified by this symbol.
• If the mattress support has been folded up, secure it against
unintentional unfolding by suitable methods.
Always carry out a performance test after any
assembly work!
10
Issued: 25.07.2007
Casa
Med Classic FS
Safety information

Advertisement

Table of Contents
loading