Inserting the mattress support motor with control unit
Open the quick-release bolt retainer (1).
Insertion direction = the mattress
support motor with control unit is fixed
under the mattress support with the motor
housing (2) pointing towards the foot end.
Caution!
• Ensure that the mattress support
motor with control unit is aligned
correctly (label on nursing bed).
Pass the receptacle of the motor side (2)
between the mattress support frame tabs
(3).
Insert the quick-release bolt (4) and
secure by closing the retainer.
Pass the second receptacle (5) between
the head end adjustment frame tabs (6).
Insert the quick-release bolt (7) and
secure by closing the retainer.