Days Casa Med Classic FS Operating Manual page 23

With dewert drive systems. nursing beds
Table of Contents

Advertisement

NURSING BEDS
Plug the electrical component plugs
into the jack sockets as shown in the
following diagram.
The labels on the plug arrangement shown
in the photo correspond to those on the
control unit.
Plug arrangement:
1 = Not in use
2 = Hand set
3 = Mattress support motor
(pre-fitted on the control unit)
4 = mattress support motor for foot
section adjustment
5 = Cable for lift motor; foot section
6 = Cable for lift motor; head section
7 = Accumulator box
(only nursing beds with
Trendelenburg's position)
*NOTE
Jack sockets not in use must be
closed using dummy plugs.
Insert the tension protection (2).
To do this:
1. Push the pegs into the drill holes.
2. Engage the clips into the recesses.
Setting up the nursing bed
Casa
Med Classic FS
Plug arrangement
Insert tension protection
Issued: 25.07.2007
23