9.0 Operating the nursing bed

Accident hazard!
• Do not use electrical medical appliances on patients in these
nursing beds.
• Do not use the nursing bed in the vicinity of strong
electromagnetic fields (for example major medical equipment).
Trapping hazard!
There is a possible danger of trapping at all joints on the nursing
bed and at support points for the mattress support.
• If the mattress support has been folded up, secure it against
unintentional unfolding by suitable methods.
When carrying out adjustments, please observe the following
points:
• Before adjusting the bed height, ensure that nothing is blocking the
full adjustment movement.
• Avoid operating the actuator motors continuously for more than two
minutes (intermittent operation).
Maximum continuous operation of actuator motors = 2 minutes.
Minimum pause after 2 minutes operation = 18 minutes.
• Do not operate more than two actuator motors at once.
• Always move the nursing bed to the lowest height when
unsupervised or for entering or exiting the bed.
• When carrying out adjustments, observe the patient in the
bed and the bed's surroundings.