Care And Maintenance; Cleaning The Measuring Instrument; Replacing Sensors - TESTO 340 Instruction Manual

Flue gas analyser
Hide thumbs Also See for 340:
Table of Contents

Advertisement

I.

Care and maintenance

This chapter describes the steps and action required in order to keep the product func-
tioning properly.
See also Regular care, p. 18.
I.1

Cleaning the measuring instrument

If the housing of the instrument is dirty, clean it with a damp cloth.
Use distilled water, or alternatively mild solvents such as isopropanol to clean the flue
gas analyzer. If using isopropanol, please refer to the instruction leaflet for the pro-
duct. Isopropanol fumes have a slight narcotic effect, and typically cause irritation of
the eyes and sensitive mucous membranes. When using it, please ensure that there
is adequate ventilation.
Do not store any objects that have come into contact with solvents and/or degrea-
sers (e.g. isopropanol) in the case. Evaporating or leaking solvents and/or degreasers
may cause damage to the instrument and to the sensors.
The use of strong or harsh alcohol or brake cleaner can result in damage to the
instrument.
I.2

Replacing sensors

A slot bridge (0192 1552) must be inserted in slots which do not have a sensor. Used
sensors must be disposed of as special waste!
The latest device software must be installed on the measuring device, see chapter
Updating device software
The measuring instrument must be switched off and the mains unit disconnected from
the mains supply.
1 Place the measuring instrument on its front.
2 Loosen screws with a screwdriver, release clip in the direction of the arrow, and
remove service cover.
3 Pull tube connections from the faulty sensor/bridge.
4 Remove the faulty sensor/bridge from the slot.
47

Advertisement

Table of Contents
loading

Table of Contents