Reflective Sensor Heads: T Mode Functions
Automatic Teaching
For automatic teaching, measurements are performed while the RIGHT
Key and the ENT Key are pressed at the same time. The center value
between maximum and minimum values is set as a threshold.
The threshold value is set when the keys are released.
The threshold can be set according to the sensing object.
Teaching
started.
Threshold:
(Maximum value - Minimum value)/2
3-4-2

Inputting Threshold Values Directly

Threshold values can be directly input into the sub-display.
Note: Generally, any value can be input. The judgement outputs, however, will
not operate for thresholds that are outside the measurement range. Also,
the decimal point cannot be changed.
Teaching
completed.
Sensor Head
Maximum value
Example: Automatic Teaching
If an error occurs when inputting a threshold value, refer
to 4-4-2 Inputting Threshold Values Directly and 5-2-3
Unable to Set Threshold Values.
Minimum value
Section 3-4
Sensor Head
Sensing object