Hardware Functions
3-2-3

Performance

Linearity
The linearity indicates how much linearity is maintained by the linear out-
put against the displacement amount (incident level). The linearity is evalu-
ated as the percentage of full scale (FS) represented by the deviation from
an ideal straight line.
For example, the 80
µ
160-
m error, or a linearity of 0.2% FS.
Resolution
The resolution is the width of the deviations in the linear output. Width of
deviation for the linear output is evaluated at
Temperature Characteristic
The temperature characteristic is measured as the deviation in the linear
output against ambient temperature changes.
The temperature characteristic is evaluated as the percentage of change
against FS for a 1°C temperature change (unit: %FS/°C).
Current Consumption
The current consumption is the maximum current consumed by the prod-
uct. Use it as a guideline to supply electric current and power.
In addition, if the product's current consumption is greater than or equal to
the specified current consumption, an error may have occurred in the prod-
uct. Immediately replace or repair it.
Ambient Operating Temperature
The ambient operating temperature is the temperature range for which
specifications are given.
Ambient Operating Humidity
The ambient operating humidity is the humidity range for which specifica-
tions are given.
Dielectric Strength
The dielectric strength is the voltage the product can endure when voltage
is applied between the product case and the charged parts.
Vibration Resistance
The vibration resistance is the vibration level that can be applied to the
product without affecting continued normal product operation.
Degree of Protection
The degree of protection indicates the tolerance of the product against
dust and water. An "IP50" degree of protection indicates the following:
• The amount of dust that may enter the product will not interfere with nor-
mal equipment operation and will not adversely affect safety.
• No special water protection is provided.
±
40-mm measurement range of the ZX-LD100 has a