Adjusting Compressibility - Agilent Technologies G1323B User Manual

1200 series handheld control module
Table of Contents

Advertisement

3
Using the Pump

Adjusting Compressibility

Adjusting Compressibility

Liquid chromatographic solvents are compressible under pressure. This
solvent compressibility produces a change in flow rate as the pressure
changes. To compensate for this effect, you can use the pump compressibility
setting to maintain a uniform and accurate flow regardless of the system
pressure. The compressibility setting can be accessed by using the Pump
Settings button [F1] in the Analysis screen.
Table 3
Table 3
Solvent
Acetone
Acetonitrile
Benzene
Carbon tetrachloride
Chloroform
Cyclohexane
Ethanol
Ethyl acetate
Heptane
Hexane
Isobutanol
Isopropanol
Methanol
1-Propanol
Toluene
Water
66
lists compressibility values for common solvents used in LC.
Compressibility Values for Common LC Solvents