System Screen - Agilent Technologies G1323B User Manual

1200 series handheld control module
Table of Contents

Advertisement

System Screen

The System screen provides access to non-routine settings and provides error
checking, control, testing and tracking usage for LC system and modules.
This screen displays the logbook together with the LC system and module
status. The EMF (Early Maintenance Feedback) icon will flash to indicate
when an EMF limit is exceeded
Current status
Figure 8
The screen provides access to:
• LC system and module controls, for example, pump on/off, lamp on/off,
• LC system and module configuration,
• module tests, and
• LC system and module records.
1200 Series Handheld Control Module Users's Guide
On/off indicators of other
modules
System Screen Information
heater on/off and injector reset,
The Agilent 1200 Series Control Module
System logbook
shows all messages from all
modules with date and time
Control Module Software
1
27

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1323a

Table of Contents