Resetting the Baseline
You can reset the baseline using the Balance button [F1] (Zero button [F1]
when using the Refractive Index Detector) in the Detector Control screen.
Please note that this procedure is not available for the Fluorescence Light
Detector:
1 Select the Control button [F1] in the System screen.
2 Select the Detector from the menu.
3 Press Enter.
4 Select the Balance button [F1].
You can achieve the same using the m (menu) key in the Detector Settings
window available from the Analysis screen.
1200 Series Handheld Control Module Users's Guide
Using the Detectors
Resetting the Baseline