Adjusting Compressibility

Adjusting Compressibility
Liquid chromatographic solvents are compressible under pressure. This solvent
compressibility produces a change in flow rate as the pressure changes. To
compensate for this effect, you can use the pump compressibility setting to maintain
a uniform and accurate flow regardless of the system pressure. The compressibility
setting can be accessed by using the Pump Settings button [F1] in the Analysis
screen.
Table 1 lists compressibility values for common solvents used in LC.
Compressibility Values for Common LC Solvents
Solvent
Compressibility (10
Acetone
126
Acetonitrile
115
Benzene
95
Carbon tetrachloride 110
Chloroform
100
Cyclohexane
118
Ethanol
114
Ethyl acetate
104
Heptane
120
Hexane
150
Isobutanol
100
Isopropanol
100
Methanol
120
1-Propanol
100
Toluene
87
Water
46
66
-6
per bar)