s1
Installation Note for Solvent
Selection Valve Upgrade Kit
(G1360A)
This note describes how to install the solvent selection valve upgrade kit into
an Agilent 1100 Series binary pump.
General Information
The programmable solvent selection valve combines two out of four solvents
for binary pump formation or selects a different solvent for flushing the
column. It cannot be used as an gradient valve. When installed the solvent
selection valve can be activated from the set-up screens of your user
interface. For channel A you can select between A1 and A2, for channel B you
can select between B1 and B2. Position 1 is the upper connection of the
valve, whereas position 2 is lower position of the valve. The valve outlet is the
middle position of the solvent selection valve.
 Copyright Agilent Technologies 2000
1

Advertisement

loading

Summary of Contents for Agilent Technologies G1360A

  • Page 1 B1 and B2. Position 1 is the upper connection of the valve, whereas position 2 is lower position of the valve. The valve outlet is the middle position of the solvent selection valve.  Copyright Agilent Technologies 2000...
  • Page 2: Delivery Checklist

    Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Delivery Checklist Ensure all parts and materials have been delivered with the upgrade kit. The delivery checklist is shown in Table 1. Please report missing or damaged parts to your local Agilent Technologies sales and service office.
  • Page 3: Safety Information

    Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Safety Information Safety Information WAR NI N G The following procedures require opening the main cover of the binary pump. Always make sure the binary pump is disconnected from the line power when the main cover is removed.
  • Page 4: Installing The Solvent Selection Valve Upgrade Kit

    Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Installing the Solvent Selection Valve Upgrade Kit Installing the Solvent Selection Valve Upgrade Kit Screwdriver Pozidriv #1 Wrench 1/4 inch Wrench 14 mm Wrench 7 mm Tools required Wrench 5 mm...
  • Page 5 Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Installing the Solvent Selection Valve Upgrade Kit 3 Disconnect the connectors of the active inlet 4 Disconnect both pumps from the main board valve and the damper from the main board...
  • Page 6 Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Installing the Solvent Selection Valve Upgrade Kit 7 Push the connector plate of the SSV cable 8 Place the cable in the slit of the lower side of assembly into the slit of the bottom foam.
  • Page 7 Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Installing the Solvent Selection Valve Upgrade Kit 11 Slide the SSV block with the two selection 12 Replace the two pump units and connect to valve into its position and fix with the two the main board (J16, J 17, J12, J27).
  • Page 8 Installation Note for Solvent Selection Valve Upgrade Kit (G1360A) Installing the Solvent Selection Valve Upgrade Kit 15 Connect the two connecting tubes between 16 Replace top foam, metal plate and top cover. the middle positions of the SSV’s and the Reinstall the leak funnel with the waste tube active inlet valve.