Single-Level Calibration Sequences
The following procedure describes how to set up a calibration sequence for an
analysis which uses single-level calibration.
There is one calibration standard (C) and 9 samples (S).
The analysis requires that:
• each sample is analyzed in duplicate,
• the calibration standard is analyzed once before the samples and
• the calibration standard is located in position 90 and the 9 sample vials are
• the method called METH1 is used for the samples and standards.
The example assumes that the sequence table is empty. If the sequence table is
not empty, use the All Lines item, accessed by pressing the Delete button [F6].
1 Select the Sequence button [F4] in the Analysis screen.
2 For sequence line 1, select the Insert button [F7] and enter:
3 Select the Calibration button [F1] to display the Calibration Settings screen for
1200 Series Handheld Control Module Users's Guide
re-analyzed once after every 2 samples,
C
S S
C
S S
in positions 1 to 9 of the autosampler tray, and
Vial Range
1 to 9
#Inj.
2
Inj. Volume
Default
Method
METH1
sequence line 1 and choose Edit.
Single-Level Calibration Sequences
C
S S
C
S S
Running Multiple-Vial Analyses
C
S
C
11
137